Fluke | 87-5

Back to list

Fluke

Digital TRMS multimeter

Code :
FLU87V
Brand :
Fluke
Serie :
80, 87-V, 87 V, 87 5, 80-V, 80V, 80 V, 80-5, 87V
Manufacturer Ref. :
87-5
Order Ref. :
Fluke-87-5
Fluke 87-V digital multimeter.
TRMS.
AC/DC voltage: 1000 V.
AC/DC current: 10 A.
Capacitance, resistance, frequency, duty cycle, temperature, diode test, continuity.
Backlight display, Bargraph.
Min/Max, AutoHold.
Refer to a friend
Share |

578,00

Add to cartOrder/Quotation
Demande de devis
  • J’ajoute mes produits à mon panier
  • J'accède à mon panier
  • Je demande mon devis à l'étape n°4
Express Quote for this product
Top
Data logger on PC : No
Functions : Large display
Backlight
Bargraph
Auto-Hold
Fuse protection
Direct input for current probe : No
ATEX : No
TRMS : Yes
Current : AC (Alternative)
DC (Continuous)
Voltage : AC (Alternative)
DC (Continuous)
Number of points (cts) : 20,000 counts
Accuracy (%) : 0.05%
Voltage resolution (V) : 10 uV
Minimum DC voltage (V) : nc
Minimum AC voltage (V) : nc
Maximum DC voltage (V) : 1000V DC
Maximum AC voltage (V) : 1000V AC
Minimum DC current (A) : nc
Minimum AC current (A) : nc
Maximum DC current (A) : 10A DC
Maximum AC current (A) : 10A AC
Measures (Type) : Resistance
Capacitance
Frequency
Temperature
Continuity test
Diode test
Min/Max
Duty cycle
Multimeter bandwidth (Hz) : 20kHz
Autonomy (Hour) : 400 hours
Safety Level : CAT I
CAT II
CAT III
Safety Overvoltage : CAT III 1000V / CAT IV 600V
Communication interface (Type) : na
Additional Characteristics : -
Package includes : TL75 test leads, AC72 alligator clips, holster, 80BK temperature probe (87V only), 9 V battery (installed), CD-ROM (user’s manual and technical notes) and operator’s guide.
Dimension (HxWxD) (mm) : 200x95x48mm
Net Weight (kg) : 600 g
Standard Manufacturer Warranty (Year) : Limited LIFETIME
Top
Fluke 80 Series V Digital Multimeters
Specifications
Voltage DC
Maximum Voltage: 1000V
Accuracy: Fluke 83 V: ±(0.1%+1)
  Fluke 87 V: ±(0.05%+1)
Maximum Resolution: Fluke 83 V: 100 µV
  Fluke 87 V: 10 µV
Voltage AC
Maximum Voltage: 1000V
Accuracy: Fluke 83 V: ±(0.5%+2)
  Fluke 87 V: ±(0.7%+2) True RMS
AC Bandwidth Fluke 83 V: 5kHz
  Fluke 87 V: 20kHz*
  *with low pass filter; 3db @ 1kHz
Maximum Resolution: 0.1 mV
Current DC
Maximum Amps: 10A (20 A for 30 seconds maximum)
Amps Accuracy: Fluke 83 V: ±(0.4%+2)
  Fluke 87 V: ±(0.2%+2)
Maximum Resolution: Fluke 83 V: 0.01 mA
  Fluke 87 V: 0.01 µA
Current AC
Maximum Amps: 10A (20 A for 30 seconds maximum)
Amps Accuracy: Fluke 83 V: ±(1.2%+2)
  Fluke 87 V: ±(1.0%+2) True RMS
Maximum Resolution: 0.1 µA
Resistance
Maximum Resistance: 50 MΩ
Accuracy: Fluke 83 V: ±(0.4%+1)
  Fluke 87 V: ±(0.2%+1)
Maximum Resolution: 0.1Ω
Capacitance
Maximum Capacitance 9,999 µF
Accuracy ±(1%+2)
Maximum Resolution 0.01 nF
Frequency
Maximum Frequency 200 kHz
Accuracy ±(0.005%+1)
Maximum Resolution 0.01 Hz
Duty Cycle
Maximum Duty Cycle 99.9%
Accuracy ±(0.2% per khz +0.1%)
Maximum Resolution 0.1%
Temperature Measurement
Fluke 87 V, 87 V/E: -200.0°C - 1090°C
  -328.0°F - 1994.0°F
  excluding probe
80 BK Temperature Probe
  -40.0°C - 260°C
  -40.0°F - 500°F, 2.2°C or 2% whichever is greater
Conductance
Maximum Conductance 60.00 nS
Accuracy ±(1.0%+10)
Maximum Resolution 0.01 nS
Diode
  Range 3 V
Resolution: 1 mV
Accuracy: ± (2 % + 1)
Duty Cycle Range
Accuracy: Within ± (0.2% per kHz + 0.1 %)

Environmental Specifications
Operating Temperature
  -20°C to +55°C
Storage Temperature
  -40°C to +60°C
Humidity (Without Condensation)
  0% - 90% (0°C - 35°C)
  0% - 70% (35°C - 55°C)
Operating Altitude
  2000 m

Safety Specifications
Overvoltage category
  EN 61010-1 to 1000 V CAT III, 600V CAT IV
Agency Approvals
  UL, CSA, TÜV, VDE listed

Mechanical & General Specifications
Size
  201 x 98 x 52 mm (with holster)
Weight
  355 g
Weigh With Holster and Flex-Stand
  624 g
Display
Digital: 6000 counts updates 4/sec.
  (Model 87 also has 19,999 counts in high-resolution mode.)
Analog: 32 segments, updates 40/sec.
Frequency: 19,999 counts, updates 3/sec at > 10 Hz.
Warranty
  Lifetime
Battery Life
Alkaline: ~400 hours typical, without backlight
Shock
  1 Meter drop per IEC 61010-1:2001
Vibration
  Per MIL-PRF-28800 for a Class 2 instrument

Top