Fluke | 114

Back to list

Fluke

TRMS Handheld digital multimeter; 6,000 counts; 0.5% accuracy

Code :
FLU114
Brand :
Fluke
Serie :
110
Manufacturer Ref. :
114
Order Ref. :
Fluke-114 EUR
Electrical multimeter for electrical troubleshooting and straightforward "go/no-go" testing in residential/commercial electrical testing.
* True RMS measurements
* 0.5% basic accuracy
* 6000 count resolution
* AutoVolt: automatic ac/dc voltage selection
* Large white LED backlight to work in poorly lit areas
* Resistance and continuity
* Min/Max/Average to record signal fluctuations
* CAT III 600 V safety rated
Refer to a friend
Share |
Top
435-Data logger on PC : No
436-Functions : Large display
Backlight
Bargraph
449-Direct input for current probe : No
27- Resolution in nb points : 6,000 counts
173-ATEX : No
424-TRMS : Yes
463-Voltage : AC (Alternative)
DC (Continuous)
45-PC interface : na
24-Accuracy (%) : 0.5%
19-Min. DC voltage-asuppr : nc
21-Min. AC voltage-asuppr : nc
37-Max. DC voltage : 600V DC
38-Max. AC voltage : 600V AC
20-Min. DC current-asuppr : na
22-Min. AC current-asuppr : na
39-Max. DC current : na
40-Max. AC current : na
33-additional measurements : Resistance
Continuity test
Min/Max
23-Voltage resolution : 0.1 mV
34-Bandwidth : 50 kHz
25-Autonomy : 400 hour
119-Safety Level : CAT I
CAT II
CAT III
120-Safety Overvoltage : CAT III 600V
36-Additional Characteristics : -
132-Package includes : Holster, installed 9 V battery, test leads, user manual
35-Dimension (HxWxD) (mm) : 167x84x46 mm
13-Net Weight : 550 g
41-Standard Manufacturer Warranty : 3 YEARS
Top

Model 114

 


The 114 is ideal for electrical troubleshooting and straightforward ‘go/no-go’ testing in residential/commercial electrical testing. It has all the basic functions plus a special feature to prevent false readings caused by ghost voltage.

Specifications
Maximum voltage between any terminal and earth ground

600 V
Surge protection

6 kV peak per IEC 61010-1 600 VCAT III, Pollution Degree 2
Display
Digital: 6,000 counts, updates 4 per second
Bar graph

33 segments, updates 32 per second
Operating Temperature

-10 °C to + 50 °C
Storage Temperature

-40 °C to + 60 °C
Battery type

9 volt Alkaline, NEDA 1604A/ IEC 6LR61
Battery Life

400 hours typical, without backlight

Accuracy Specifications
DC millivolts
Range: 600.0 mV
Resolution: 0.1 mV
Accuracy: ± ([% of reading] + [counts]): 0.5% + 2
DC volts
Range/Resolution: 6.000 V / 0.001 V
Range/Resolution: 60.00 V / 0.01 V
Range/Resolution: 600.00 V / 0.1 V
Accuracy: ± ([% of reading] + [counts]): 0.5% + 2
Auto volts
Range: 600.0 V
Resolution: 0.1 V
Accuracy: 2.0 % + 3 (dc, 45 Hz to 500 Hz)4.0 % + 3 (500 Hz to 1 kHz)
AC millivolts1 true-rms
Range: 600.0 mV
Resolution: 0.1 mV
Accuracy: 1.0 % + 3 (dc, 45 Hz to 500 Hz)2.0 % + 3 (500 Hz to 1 kHz)
AC volts1 true-rms
Range/Resolution: 6.000 V / 0.001 V
Range/Resolution: 60.00 V / 0.01 V
Range/Resolution: 600.0 V / 0.1 V
Accuracy: 1.0 % + 3 (dc, 45 Hz to 500 Hz)2.0 % + 3 (500 Hz to 1 kHz)
Continuity
Range: 600 Ω
Resolution: 1 Ω
Accuracy: Beeper on < 20 Ω, off > 250 Ω; detectsopens or shorts of 500 μs or longer.
Ohms
Range/Resolution: 600.0 Ω / 0.1 Ω
Range/Resolution: 6.000 kΩ / 0.001 kΩ
Range/Resolution: 60.00 kΩ / 0.01 kΩ
Range/Resolution: 600.0 kΩ / 0.1 kΩ
Range/Resolution: 6.000 MΩ / 0.001 MΩ
Accuracy: 0.9 % + 1
Range/Resolution: 40.00 MΩ / 0.01 MΩ
Accuracy: 1.5 % + 2

Notes:
1) All ac voltage ranges exept Auto - V / Lo-Z are specified from 1% to 100% of range. Auto - V / Lo-Z is specified from 0.0V.
2) Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage.

Mechanical & General Specifications
Size
Height: 16.7 cm (6.57")
Length: 8.4 cm (3.31")
Width: 4.62 cm (1.82")
Weight

392 g (0.866 lbs)
Warranty

3 years

Top