_AlertIE

Tektronix | TDP1500

Sonde différentielle haute tension, 1.5GHz, 200 kOhm, +/-25V, x1/x10

Tektronix
Code Testoon : TEKTDP1500
Référence commande : TDP1500
Marque : Tektronix
Sonde différentielle haute tension, 1.5GHz, 200 kOhm, +/-25V
Temps de montée: <700 ps
Capacité d'entrée : <265 pF
Résistance d'entrée : 200 kOhm
Tension d'entrée en mode commun : ±7 V (1X); ±7 V (10X)
Gamme de tension d'entrée différentiel : ±8.5 V (10X); ±850 mV (1X)
TTCHT
Quantité
Expédié sous 3 semaines
Devis EXPRESS pour cet article
Devis EXPRESS pour cet article
Description
TDP1500
Applications
Design. validation. debugging. and characterization of common highspeed serial bus designs:
 - I 2C
 - CAN/LIN
 - SPI
 - Serial ATA
 - Ethernet (GbE)
 - USB 2.0
 - FIreWire (1394b)
Signal integrity. jitter. and timing analysis
Manufacturing. engineering. and test
High-voltage differential probes
Specifically designed for use and direct connection to oscilloscopes with the TekVPI™ probe interface. the TDP1500 and TDP3500 Differential probes achieve high-speed signal acquisition and measurement fidelity by solving three traditional problems: 
 - DUT loading effects - Are reduced by lower input capacitance and high input resistance
 - DUT connectivity - A variety of accessories exist for attaching to small SMDs. some come standard or recommended
 - Maximizing of system (oscilloscope and probe) bandwidth - Probing solutions for all measurements for TekVPI interface oscilloscope models up to 3.5 GHz
Key performance specifications
1.5 GHz (TDP1500) probe bandwidth
≤1 pF (TDP1500) ) differential input capacitance
200 kΩ (TDP1500)  differential input resistance
±25 V (TDP1500) DC + pk AC input voltage
>60 dB at 1 MHz and >25 dB at 1 GHz CMRR
Key features
Outstanding electrical performance
 - 3.5 GHz and 1.5 GHz bandwidth models - accurate measurements for serial and digital applications
 - Excellent common mode rejection – reduces measurement errors in higher common environments
 - Low capacitive and resistive loading – maintains signal fidelity and reduces DC biasing interactions
Versatile mechanical performance
 - Compact probe head size for probing small geometry circuit elements
 - DUT attachment accessories enable connection to fine-spaced SMDs
 - Robust design for reliability
Easy to use
 - Connects directly to oscilloscopes with the TekVPI™ probe interface
 - Provides automatic units scaling and readout on the oscilloscopes display
 - Easy access to probe comp box controls or oscilloscope probe menu display for probe status. setup control. and diagnostic information
Integrated Scope/Probe system
 - Direct connection to and powered from the TekVPI™ oscilloscope interface (Connects directly to TekVPI scopes without the need of an external power supply. like many competitors require)
 - Single-button oscilloscope probe menu access
 - Setup and control from probe comp box or oscilloscope user interface
 - AutoZero - zeros out output offset
 - Remote GPIB/USB probe control through the oscilloscope
Specifications 
Attenuation (probe only): x1; x10
Bandwidth (probe only): ≥1.5 GHz
Rise time (probe only): <265 ps="" div="">
Input capacitance: <1 pf="" div="">
Input resistance: 200 kΩ
Propogation delay: 5.4 ns
Common mode input voltage: ±7 V (1X); ±7 V (10X)
Differential input offset range: ±8.5 V (10X); ±850 mV (1X)
Spécifications
Garantie Constructeur Standard 1 AN
017-Capacité d'entrée (pF) 1 pF
380-Tension différentielle d'entrée max. (V) 8.5 V
048-Temps de montée (s) 265 ps
015-Impédance d'entrée (Ohm) 200 kOhm
004-Bande Passante (Hz) 1.5 GHz
012-Attenuation sonde (X) 1X | 10X
379-Tension différentielle d'entrée min. (V) 850 mV
052-Longueur de câble 1.2 m
120-Surtension de Sécurité nc
132-Livré avec Manuel d'utilisation, sacoche en nylon, kit d'accessoires TDP1500, certificat de traçabilité, documentation TDP1500
035-Dimensions (HxLxP) (mm) nc
013-Poids Net 160 g