<< Return

|
 |
Testoon Price : 344 € Exc VAT Quantity :


Brand : TTI - Thurlby Thandar Instruments
Manufacturer Ref. : TGP110
Presentation

Pulse generator with Delay.
0.1Hz to 10MHz frequency range
Independent control of pulse frequency, width and delay
50ns minimum pulse width
Squarewave, double pulse & delayed pulse modes
Free-run, gated and triggered modes
50 Ohm output: 0.1V to 10V amplitude
TTL/CMOS and Sync outputs
Low-cost
<< Similar products

Download the data sheet
|
| |
 |
| Frequency Range (Hz) : | 10 MHz
| | Delay range (s) : | 50ns - 5s
| | Pulse width min (s) : | 50ns
| | Pulse width max (s) : | 5s
| | Max output voltage (V) : | 10Vpp
| | Wobulation (Y/N) : | No
| | Additional output (TTL and CMOS) : | TTL
| | Waveforms (Type) : | Square Pulse
| | Safety Overvoltage : | nc
| | Display/resolution : | No display
| | Additional Characteristics : | -
| | Package includes : | nc
| | Dimension (HxWxD) (mm) : | 140x220x230 mm
| | Net Weight (kg) : | 1.6 kg
| | Standard Manufacturer Warranty (Year) : | 1 YEAR | |
|
| |
 |
Brand : TTI - Thurlby Thandar Instruments Manufacturer Ref. : TGP110 Weight : 2.6 kg
Thurlby Thandar TGP110
An essential instrument
The generation of pulses for the stimulus and control of electronic
systems is beyond the capability of all but the most sophisticated of
function generators. The architecture of a dedicated pulse generator
enables it to generate pulses of a set width regardless of the repetition
rate offering duty cycles which can extend down to 1 in 100 million.
In addition to continuous operation, single or multiple pulses can be
generated in response to trigger or gating signals with precisely defined
timing relationships. A dedicated pulse-waveform output amplifier provides
flat top pulses with fast rise and fall times at variable amplitude.
Variable delay
The TGP110 offers selectable delay between the sync output and pulse
output. In triggered mode this also sets the delay between a trigger
signal and the start of the pulse.
Wide pulse range
The TGP110 can generate pulse widths in the range 50ns to 5s. There
are eight overlapping decade ranges with vernier control within each
range.
The period range is 100ns to 10s, equivalent to a repetition rate range of
10MHz to 0.1Hz. Delay is independently adjustable over the same range as
pulse width. A complement switch inverts the mark-space polarity.
Squarewave and double pulse
In square mode, squarewaves are generated at a frequency set by the
period controls alone. This provides a convenient means of generating
variable period edges where the pulse width is unimportant, for example.
In double pulse mode, a second pulse is generated within every period at a
set delay after the start of the first pulse. The delay is independently
adjustable.
| SPECIFICATIONS |
| PERIOD, PULSE
WIDTH, DELAY |
| Each parameter is
variable within 8 overlapping decade ranges with a vernier providing
continuously variable control within each range. |
| PERIOD |
| Range: |
100nsec to 10sec (10MHz to 0·1Hz). |
| Jitter: |
<0.1%. |
| PULSE
WIDTH |
| Range: |
50nsec to 5sec |
| Jitter: |
<0.1%. |
| DELAY |
| Range: |
50nsec to 5sec |
| TRIGGER,
GATE |
| RUN |
| Normal operational
mode in which pulses are generated continuously at 0.1Hz to 10MHz. |
| TRIGGERED |
| DC to 10MHz pulse
train in synchronism with external trigger pulses; pulse width
determined by pulse width controls. Trigger can be generated manually
from front panel button. |
| GATED |
| 0·1Hz to 10MHz
pulse train, parameters set by period and pulse width controls, starts
synchronously with leading edge of gate input. Last pulse is completed
at the end of gating period. Gating signal can be generated manually
from front panel button. |
| PULSE
MODES |
| NORMAL PULSE |
| One pulse is
generated each period. The delay setting is ignored. |
| SQUAREWAVE |
| 0·1Hz to 10MHz
squarewave, frequency set by the period controls. Pulse width and delay
settings ignored. |
| Mark : Space ratio: |
1:1 ±10%. |
| DOUBLE
PULSE |
| A second pulse is
generated after a delay set by the delay controls; the delay is related
to the leading edge of the first pulse. |
| DELAYED
PULSE |
| A pulse is
generated after a delay set by the delay controls; the delay is related
to rising edge of the trigger signal. |
| INPUTS |
| GATE/TRIG INPUT |
| Frequency range: |
DC - 10MHz |
| Signal range: |
TTL threshold; max. input ±10V. |
| Min. pulse width: |
>30nsec. |
| Input Impedance: |
Typically 10kW. |
| OUTPUTS |
| 50W
OUTPUT |
| Amplitude: |
Two switch selectable ranges
of 0·1V - 1·0V and 1V - 10V from 50W.
(50mV to 500mV and 500mV to 5V into 50W).
Adjustable within ranges by a single turn vernier. |
| Rise/Fall Times: |
Typically 10nsec into 50Wload.
Maximum 15ns. |
| Aberrations: |
Typically <5% for output
set at >20% of range maximum, into 50W |
| AUX
OUTPUT |
| Duplicates 50W
output but at a fixed CMOS/TTL level. |
| SYNC
OUTPUT |
| Amplitude: |
A positive going pulse at CMOS/TTL
level. |
| Timing: |
Leading edge starts >20nsec
before the TTL/50WW output
transition. |
| Duration: |
Typically 30nsec. |
| COMPLEMENT
SWITCH |
| Inverts the AUX
and 50W
outputs. |
| GENERAL |
| Power: |
230V or 115V AC nominal
50/60Hz, adjustable internally; operating range ±14% of nominal; 20VA
max. |
| Size: |
140 x 220 x 230mm (HxWxD) |
| Weight: |
1.6kg (3.5lb) |
| Operating Range: |
+ 5°C to 40°C, 20-80% RH. |
| Storage Range: |
- 40°C to 70°C |
| Safety: |
Complies with EN61010-1. |
| EMC: |
Complies with EN55081-1 and
EN50082-1. |
|