Keithley 2110
Multimètre numérique de table 5.5 digits, 1000 V, 10 A, 100 MOhm, USB standard, GPIB option
Modèle avec interface USB seulement.
Pour l'interface GPIB, voir le modèle KEI2110-GPIB visible ici.
=======================
Multimètre Numérique de table modèle 2110, 1 kV/10 A/100 MOhm
5.5 chiffres de résolution
Face avant 2 ou 4 fils de mesuresa
15 fonctions de mesures incluant les mesures de capacité et de thermocouples
Une interface USB 2.0 & GPIB (2110-GPIB seulement) utilisable avec des tests de commande SCPI.
DMM 2110-240
Key Performance Specifications
5.5 Digit Resolution
Basic VDC Accuracy of up to 0.012 % (1 yr.)
100 mV to 1000 V Voltage Range. with up to 10 mV Resolution
10 mA to 10 A Current Range. with up to 100 µA Resolution
100 Ω to 100 MΩ Range. with up to 1 kΩ Resolution
CAT I 1000 V. CAT II 600 V
Available Functions and Features
Volts. Ohms. and Amps Measurements
True RMS (AC. AC + DC) Measurements
Diode and Continuity Testing
Frequency and Period Measurements
Temperature and Capacitance Measurements
2×4 Ohms 4-wire Measurement Technique
TrendPlot™ Paperless Data Recorder Mode
Measurement Statistics
Histogram Mode
Connectivity
Front 2×4 Measurement Inputs
USB Host Port on Front Panel for Easy Storage of Measurement Data and Instrument Settings
GPIB Ports on Rear Panel for Quick PC Connectivity
Includes USB Interface Adapter Cable
Includes National Instrument’s LabVIEW Driver.
APPLICATIONS
BUILT for Production Testing
Rhe model 2110 Digital Multimeter is ideal for Application in manual. semi-automatic. and automatic testing of low-cost electronic devices. circuits. modules. electrical components. and semiconductor components. Key features include:
Speed : up to 50.000 readings per second
Control: GPIB (Optional) and USB interfaces. accepting SCPI (IEEE-488.2) Commands
External BNC trigger lines
NIST traceability (with included calobration certificate)
Built for General Porpose Uses
The model 2110 Digital Multimeter is also ideal for bench uses such as research. development. service. calibration. and teaching. Bench-oriented features include:
- Accuracy: 0.012 % basic DCV accuracy
- Easy-to-operate panel
- Easy waveform plotting and data collection with KI-Tool and KI-Link
- Store up to 2000 readings
SELECTOR GUIDE
Models | 2110 | 2000 | 2010 |
Digits | 5½ | 6½ | 7½ |
Expansion Channel | N/A | 10 | 10 |
DC Volts |
|
|
|
Sensitivity | 1 µV | 1 µV | 1 nV |
Maximum Reading | 1000 V | 1000 V | 1000 V |
Basic Accuracy | 0.012% | 0.002% | 0.0018% |
AC Volts (TRMS) |
|
|
|
Sensitivity | 1 µV | 100 nV | 100 nV |
Maximum Reading | 750 V | 750 V | 750 V |
Basic Accuracy | 0.12% | 0.05% | 0.05% |
Bandwidth | 10 Hz–300 kHz | 3 Hz–300 kHz | 3 Hz–300 kHz |
dB. dBm |
| - | - |
Frequency. Period | - | - | - |
Peak/Avg/RMS | RMS |
|
|
AC. AC+ DC | AC | AC |
|
Ohms (2/4 Wire) |
|
|
|
Sensitivity | 1 MOhm | 100 µOhm | 100 µOhm |
Maximum Reading | 100 MOhm | 120 MOhm | 120 MOhm |
Basic Accuracy | 0.02% | 0.008% | 0.0032% |
DC Amps |
|
|
|
Sensitivity | 0.1 µA | 10 nA | 1 nA |
Range Span | 10 mA–10 A | 10 mA–3 A | 10 mA–3 A |
Basic Accuracy | 0.015% | 0.03% | 0.03% |
AC Amps (TRMS) |
|
|
|
Sensitivity | 10 µA | 1 µA | 1 µA |
Range Span | 1 A–10 A | 1 A–3 A | 1 A–3 A |
Basic Accuracy | 0.3% | 0.1% | 0.1% |
Bandwith | 10 Hz–5 kHz | 3 Hz–5 kHz | 3 Hz–5 kHz |
General Features |
|
|
|
Interface | USB. GPIB (opt.) | GPIB. RS-232 | GPIB. RS-232 |
Reading Hold | - | - | - |
Digital I/O | Trigger In/Out |
|
|
Reading Memory | 1024 rdg | 1024 rdg | 1024 rdg |
Maximum Speed | 50K rdg/s | 2000 rdg/s | 2000 rdg/s |
Temperature Meas. | T/C. RTD. Thermistor | T/C | T/C. RTD |
41-Standard Manufacturer Warranty | 3 YEARS |
33-additional measurements | Resistance | Capacitance | Frequency | Temperature | Continuity test | Diode test |
424-TRMS | Yes |
462-Current | AC (Alternative) | DC (Continuous) |
463-Voltage | AC (Alternative) | DC (Continuous) |
23-Voltage resolution | 10 µV |
34-Bandwidth | 300 kHz |
24-Accuracy (%) | 0.012 % |
37-Max. DC voltage | 1000 V |
38-Max. AC voltage | 750 V |
382-Resistance measurement method | 2 wires | 4 wires |
39-Max. DC current | 3 A |
40-Max. AC current | 10 A |
119-Safety Level | CAT I | CAT II |
En Stock | No |
120-Safety Overvoltage | Cat 1 1000 V et Cat II 600 V |
2-RS232 Interface | No |
172-GPIB | In option |
132-Package includes | Reference Manuel on CD, Specifications, LabVIEW Driver, Keithley I/O Layer, USB Cable, Power Cable, Safety Test Lead, KI-Tool, and KI-Link Add-in (both Microsoft Word and Excel versions), Calibration Certificate |
35-Dimension (HxWxD) (mm) | 107x252.8x305 mm |
13-Net Weight | 2.23 kg |
- Brochures- fiches techniques